The Tritium sampler 4 vials (MARC 7000) equipment is designed to sample the tritium which is contained in a volume of air (gas H3, tritiated water HTO or organically combined). Tritiated water vapor is trapped in the first two feeding bottles by means of the bubbling principle. To trap the tritium which is combined to organic materials, an oxidation reaction is created in the oven. A catalyser is used to lower the combustion level. This causes tritium to react chemically to form tritiated water vapor which is trapped in feeding bottles n°3 & 4. After a certain time, the tritiated water contained in the bottles is measured in a laboratory. The quantity of tritium measured is related to the volume of air which has passed through the equipment.
Read more about the Tritium sampler 4 vials (MARC 7000) on the SDEC website
Voor dit product is één download beschikbaar.
Vul uw e-mailadres in om toegang te krijgen tot alle downloads van PEO.
Kies uw cookievoorkeuren. Om de beste ervaringen te bieden, gebruiken we o.a. cookies om apparaatinformatie op te slaan en/of te openen. Door toestemming te geven voor deze cookies kunnen we bepaalde gegevens op deze site verwerken. Als u geen toestemming geeft, kan dit een negatief effect hebben op de werking.